G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Operation
167
Checking the limit of the suspension arm's downward movement
– The minimum working distance (height) from the surgical field has
been set using the adjustment screw for limiting downward travel.
S88 floor stand only: stand base
– Check that the locking pedal has been pressed and that the stand is
securely locked in position.
S8 ceiling mount only: grip
– Check that the working position has been set.
Check the accessories
– Using the manuals provided, check that the other equipment (surgical
microscope, coobservation tube, video system, etc.) is functioning
properly.
Foot control panel
– Check that the plug of the foot control panel has been connected.
– Check that the power switch of suspension system has been switched
on.
•
Check that all functions assigned to the respective keys on the foot
control panel are working properly.