68
Description
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Note:
When inserting a new lamp module, make sure that the knob (2) is set
to "1“. If the first bulb fails, switch to the second bulb in logical se-
quence.
3
Indicator: backup bulb is in use
When the red segment in the knob (2) lights up, the backup bulb is in
use.
4
Filter selector knob
The filter knob has two positions:
5
Counter
The counter records the service hours of the xenon light source.
•
Change the xenon bulbs after about 500 hours of operation to prevent
them from exploding. Then reset the counter to "0" by pressing reset
button (6).
6
Reset button
The reset button resets the service hour counter to "0".
7
Opening the lamp module
When you press this button, the lamp module is slightly ejected.
•
To change the bulb, pull out the lamp module as far as it will go. Turn
the knob (2) 180° until it snaps in place. This moves the backup bulb
into the illumination beam path.
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
•
Replace the xenon lamp in good time.
•
Reset the service hour counter to "0" after replacing the lamp.
0
no filter
1
Filter swung in
(No integrated filter in the standard configuration)