Maintenance / Further information
207
OPMI
®
Vario surgical microscope on suspension system
Insufficient surgical field illu-
mination.
Brightness level set too low. Adjust brightness using the
brightness control on the il-
lumination system. See
user's manual for suspen-
sion system.
–
Light guide not properly in-
serted in lamp and/or micro-
scope.
Insert light guide as far as it
will go.
Defective light guide (illumi-
nation not uniform).
Contact service dept.
–
Zoom system inoperative.
Defective motor.
Adjust magnification manu-
ally using the zoom knob.
Contact service dept.
Zoom system always moves
to upper or lower end posi-
tion.
Switch suspension system
to MANUAL mode.
Adjust magnification manu-
ally using the zoom knob.
Contact service dept.
Focusing system inopera-
tive.
Focus stop button pressed. Press the focus stop button
again. The button must not
be lit.
Defective motor.
Adjust the working distance
manually using the focusing
knob.
Contact service dept.
Focusing system always
moves to upper or lower end
position.
Switch suspension system
to MANUAL mode.
Adjust the working distance
manually using the focusing
knob.
Contact service dept.
Problem
Possible cause
Remedy
See
Problem
Possible cause
Remedy
See
Total magnification is incor-
rectly displayed.
Eyepiece magnification has
not been correctly entered.
Enter current eyepiece
magnification at EYEPIECE
in configuration menu 3.