Trouble-shooting
•
This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety
must be checked once every 12 months. We recommend having this
check performed by our service representative as part of regular main-
tenance work.
If a failure occurs which you cannot correct using the trouble-shooting
table, attach a sign to the instrument stating it is out of order and con-
tact our service representative.
OPMI
®
Vario surgical microscope
Problem
Possible cause
Remedy
See
No function at all.
Suspension system
switched off.
Switch on the suspension
system.
–
Line power failure.
Contact in-house
electrician.
–
Surgical field illumination in-
operative.
Illumination system of sus-
pension system switched
off.
Switch on illumination
system of suspension
system. See user's manual
for suspension system.
–
Defective lamp.
Change the lamp. See
user's manual for suspen-
sion system.
Light guide not properly in-
serted in lamp or micro-
scope.
Insert light guide as far as it
will go.
Lamp module in suspension
system has no contact.
Insert lamp module as far as
it will go.
–
Failure of electronics.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
–