G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Maintenance / Further information
213
Video monitor
No surgical field illumina-
tion.
Xenon lamp does not ignite. Switch off illumination on
suspension system.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
--
Xenon lamp is lit, but beep
sounds intermittently.
Defective lamp control
system.
Switch off illumination on
suspension system.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
--
Problem
Possible cause
Remedy
See
Problem
Possible cause
Remedy
See
No imge display.
Main switch is switched off. Switch on the main switch.
-
Power cable not correctly
plugged in.
Plug in the power cable.
-
Displayed message
"NO SIGNAL"
No signal source con-
nected.
Connect a signal source.
-
No camera connected.
Connect a camera.
-
Connected video cable is
defective.
Connect an intact video
cable.
-
Image is not centered.
A signal source is present
on the VGA port, and no au-
tomatic image adjustment
has been performed.
Perform automatic image
adjustment.
Image is too dark or too
bright.
A signal source is present
on the VGA port, and no au-
tomatic image adjustment
has been performed.
Perform automatic image
adjustment.