Failure of the focusing function
•
Adjust the focus manually using focusing knob (1).
•
If the focus drive moves into an end position, press the "MANUAL"
button on the control panel of your suspension system. The "MANU-
AL" key permits you to switch to manual operation. The motorized
functions of the surgical microscope are deactivated.
– Continue surgery by manually operating the suspension system
and the surgical microscope.
or
– Manually focus the surgical microscope by moving the suspension
arm or using the lifting column of the suspension system.