IOLMaster 700
3.2.7
Shutter (PCB & Cable)
Precondition:
Shut down and switch off the instrument
The measurement head side covers have been removed from the IOLMaster 700.
The side dust cover has been removed.
Procedure:
Caution!
When disconnecting the connectors be
careful to not accidentally
damage/remove the socket.
1)
Disconnect the connector at main control
PCB.
2)
Remove two screws (Torx 10) on the fixation
points.
Fig. 26 – Shutter overview
Reassembly:
1)
Reassemble in the reverse order.
2)
Perform a system checkout in accordance
with the instrument checklist.
Fig. 27 – Shutter
Screws
Service Manual
Confidential and Proprietary
SM-30-4060
Order No. 2075-325
3 - 12