3.2
Measurement Head
3.2.1
Measurement Head Side Covers
Precondition:
The instrument has been shut down and switched off.
Procedure:
1)
Remove the front cover retaining ring by
turning in an anticlockwise direction.
See Fig. 1
.
Fig. 1 – Front cover ring
2)
To remove the left side cover (viewed from
patient side) insert a 2mm hex key into the
three apertures as shown in
Fig. 2
. Turn
anticlockwise to open the fastener
(DON’T OVERTURN – refer to Step 3 for
detailed information).
Fig. 2 - measurement head side covers
Apertures
SM-30-4060
Confidential and Proprietary
Service Manual
3 - 3
Order No. 2075-325