Fig. 55 – Start Adjust Retina Mode
As described above accurately centre the
23)
middle reflection from the test eye within the
green circle (from cross hair) displayed within
the software.
Click the “Start Adjust Retina Mode” button
24)
Caution!
Do not bump/knock/disturb the
IOLMaster 700 (or table) while the
program is running – this can lead to
inaccurate measurements.
Without moving the measurement head
25)
click the “Start Adjust AS Mode”
Caution!
Do not bump/knock/disturb the
IOLMaster 700 (or table) while the
program is running – this can lead to
inaccurate measurements.
Fig. 56 – Start Adjust AS Mode
Accurately centre
the middle
reflection (pulsing)
within the green
circle. You will
notice a line to
the right, this is
allowed outside of
the green circle.