S301 – Exchange complete measurement head
Description:
Use this function after replacement of the
complete measurement head.
The use-case covers the following tasks:
(1) Retrieve device-settings from both
measurement head and local files.
(2) Join both sets of device-settings:
Use serial-numbers of PC and device from
local files,
use testeye-settings from local files,
use all other device-settings from
measurement head (factory-adjusted).
(3) Persist the joined settings to both device-
head and local-files.
(4) Backup the joined settings onto the
service-drive (mSata).
Fig. 14 – S301 Exchange complete measurement head
(1)
Instructions: (Step 1)
Prerequisites:
–
The old measurement head has been
dismounted and replaced by a new one.
–
The new measurement head is
completely factory-adjusted.
–
PC and PI, including the test-eye, were
not changed.
–
The new measurement head is properly
connected and the firmware is running.
(This can be verified using ‘use-case’
(1) Ensure, that all prerequisites are fulfilled.
(2) Click button 'Continue' to proceed or
‘Step back’ to the use-case selection
view.
Fig. 15 – S301 Exchange complete measurement head
(2)
SM-30-4060
Confidential and Proprietary
Service Manual
4 - 9
Order No. 2075-325