SNR Confirmation:
The ‘SNR_Testeye’ + ‘SNR_Limit’ must be
1)
at least 98:
If YES – proceed to 27)
If NO – click “EXIT” and return to Step 20)
(possible alignment problem)
if Step 20) has been repeated and problems
still exist then contact
Click “Store Results to Config” (1).
2)
Click “EXIT” (2) (the test eye will be hidden
3)
and the tool will close).
Press any key to close the cmd.exe window.
4)
Fig. 57 – Confirm SNR
Run (double click) the
5)
StoreDeviceSettings.bat.
Press any key to close the cmd.exe window.
6)
Fig. 58 – Run the StoreDeviceSettings.bat
The new settings must be saved to the measurement head refer to appropriate service use case.
7)
Navigate to D:\RawCalibrationData\SNRdata
8)
9)
Perform a system checkout in accordance with the instrument checklist.
10)
Confirm
(SNR_T SNR_Limit) ≥ 98
eg. 43.8 + 56 = 99.8
1
2
SM-30-4060
Confidential and Proprietary
Service Manual
4 - 25
Order No. 2075-325