Currently only 5 pre-configured control-files are available, namely:
clean_optics_part1.control - Sets lens position for cleaning
1)
clean_optics_part2.control - Sets lens position for cleaning
2)
testeye_show.control
- Moves the Test eye out of the PI (eg. for cleaning)
3)
testeye_hide.control
- Moves the Test eye into the PI
4)
laser_ON.control
- Switches the laser ON
5)
4.3.3
Calibration use-cases
Refer to
for a detailed description about use-case principles.
M101 – Calibrate camera measurement resolution
Description:
This use-case is currently not possible in the field as hardware tools are required which aren’t
currently available.
M102 – Calibrate keratometry values
Description:
This use-case is currently not possible in the field as hardware tools are required which aren’t
currently available.