4.3
Service Tool
To carry out IOLMaster 700 service activities a
separate service tool must be used. The service
tool can only be executed from the service mode.
The passwords must not be distributed to any
non-authorised person.
The usage of the Service Tool is restricted to
certified field service engineers (FSEs), trained by
an authorised ZEISS trainer.
Fig. 6 – Service activities
4.3.1
Running the Service Tool
Enter the service mode.
Refer to section 4.2.1, see page 3.
1)
Navigate to “C:\Program Files\CZM\IOLMaster700ServiceTool\”.
2)
Double click the “Czm.Biometry.ServiceTool.exe” application.
3)
Enter the required password and press “Logon”.
4)
Attention!
When using the Service Tool the chinrest will automatically move. Ensure no patient
head resides on the chinrest when using the Service Tool!
4.3.2
Standard use-cases
The service tool is divided into “use-cases”, as further service activities become available more or less
use-cases may become available. The principle of a use-case is that each alignment/replacement
procedure will have its own use case reducing the chance of error by providing an on-screen
procedure and only interacting with hardware as required.
Currently only 2 types of use cases exist, ‘standard’ and ‘calibration’ use-cases.