Procedure (user management):
This procedure will attempt to restore a
backup of the user management database.
Begin by running
1)
“DeleteUserManagementDatabaseFile.bat”.
Restart the instrument and check whether
2)
the application software recovered from the
error.
Fig. 67 – Simple user management DB removal
Procedure (user management complete):
If a simple user management database
removal did not resolve the problem this
procedure will completely remove the user
management database.
Run
1)
“DeleteUserManagementDatabaseComplete.
bat”.
Press “r” key and “Enter” key.
2)
Restart the instrument and check whether
3)
the application software has recovered.
Fig. 68 - Complete user management DB removal
Note
After the database has been successfully restored/removed, perform a system checkout
in accordance with the instrument checklist.
SM-30-4060
Confidential and Proprietary
Service Manual
4 - 33
Order No. 2075-325