![background image](http://html1.mh-extra.com/html/zeiss/iolmaster-700/iolmaster-700_service-manual_3344775031.webp)
Parts Removal Replacement
3.3.3
Computer Module Replacement
Note
The new computer module comes with two hard drives and mSATA which must be
installed in the faulty computer module and returned to CZM.
Precondition:
The computer module and cover have been removed from the IOLMaster 700.
Procedure:
1)
Remove the cover from new computer
module (3 screws – Torx 10 + 8) and remove
the two hard drives (empty) and mSATA
refer to 3.3.8, page 19 and 3.3.7, page
2)
Remove the two hard drives and mSATA
refer to 3.3.8, page 19 and 3.3.7, page 18
from the faulty computer module and install
in the new computer module.
3)
Install the two hard drives and mSATA
delivered with the new computer module in
the faulty computer module.
4)
Note/photograph the computer serial
number. This will be required by the service
tool!
Fig. 33 – Computer Module Replacement
Reassembly:
1)
Reassemble following the reassembly steps described in
2)
It may be necessary to change the boot order.
3)
Using the IOLMaster 700 service tool carry out standard use case S401.
4)
Create a hard drive backup. See Backup.
5)
Perform a system checkout in accordance with the instrument checklist.
SM-30-4060
Confidential and Proprietary
Service Manual
Version 03, 2014-10-13
3 - 15