Prerequisites:
–
The bootloader program for the IOLMaster
700 and the new firmware file are present
and accessible at an arbitrary disk location.
–
The ServiceTool is properly configured to use
this bootloader program and its supplied
firmware file.
1) Ensure, that all prerequisites are fulfilled and
verify the configured bootloader
commandline.
2) Click button 'Continue' to proceed or 'Step
Back' to go back to the use case selection
view.
Fig. 9 – S105 Update measurement head
firmware (2)
Instructions:
(1) Wait, till the bootloader-program indicates
that the firmware-update is complete and the
IOLMaster 700 has completed the
measurement head reboot.
(2) Exit the bootloader-program.
(3) Click 'Finish' button (Returns to the use-case
selection view).
Fig. 10 – S105 Update measurement head
firmware (3)
SM-30-4060
Confidential and Proprietary
Service Manual
4 - 7
Order No. 2075-325