Run (double click) the
20)
StartPolarizationAdjustTool.bat.
Be patient as the software is copied and
21)
started.
Fig. 53 – Run the StartPolarizationAdjustTool.bat
A new window will open, ensure the chinrest
22)
is free and press “Continue”. The
software/hardware takes some time to
prepare.
Fig. 54 – The software/hardware takes to prepare