![background image](http://html1.mh-extra.com/html/zeiss/iolmaster-700/iolmaster-700_service-manual_3344775034.webp)
Parts Removal Replacement
3.3.6
Computer Module Cooling Fan
Precondition:
The computer module and cover have been removed.
Procedure:
1)
Disconnect the appropriate cooling fan cable
at CIF (cut zip tie as appropriate).
Note
To remove the left fan (
see Fig. 37
) it’s
necessary to remove the PCIe Adapter
PCB.
Fig. 37 – Cooling Fans
2)
Pull the four rubber grommets upwards and
cut off with scissors, remove excess from
under computer module.
3)
Remove cooling fan leaving the rubber holder
for the new cooling fan.
Reassembly:
1)
Reassemble in the reverse order, making sure
to cut off excess from the rubber grommet
once installed (
Fig. 38
) – 3 rings should
remain.
2)
Perform a system checkout in accordance
with the instrument checklist.
Fig. 38 – Rubber Grommets
3.3.7
mSATA (service drive)
Precondition:
The computer module and cover have been removed.
Procedure:
1)
Remove the two screws (Torx 6) (
Fig. 39
).
2)
Remove mSATA.
Reassembly:
1)
Reassemble in the reverse order, ensuring
that the sticker is facing out.
2)
Perform a system checkout in accordance
with the instrument checklist.
Fig. 39 – mSATA replacement
Fans
Screws
Service Manual
Confidential and Proprietary
SM-30-4060
3 - 18
Version 03, 2014-10-13