5 First Operating Steps | 5.6 Finding Appropriate Detector Settings
ZEISS
NOTICE
Touch alarm disabled
When beam deceleration or stage bias is activated, the stage touch alarm is automatically de-
activated. A warning is displayed to inform you that the touch alarm is not enabled. This
means that any stage collisions will not be detected.
4
Move the stage carefully.
Prerequisite
ü
Stage is initialized.
ü
HV mode is active.
ü
Standard conducting dovetail-mounted specimen holder is fitted.
Procedure
1. In the
Stage Navigation Bar
, click
Settings
.
INFO:
To open the
Stage Navigation Bar
, navigate to
View > Toolbars
and activate
Stage Navigation Bar (for Widescreen users)
. Alternatively, you can access the
Stage
Navigation Bar
via
Stage > Navigation
.
à
The
Stage Navigation Settings
dialog is displayed.
2. Activate the
Safe Navigation
checkbox.
3. In the
Stage Navigation Settings
dialog, click
Sample Holder Gallery
.
à
The
Sample Holder Gallery
dialog is displayed.
On the left hand side, a list of icons represents the specimen holders.
4. In the
Sample Holder Gallery
dialog, select the installed specimen holder.
5. Activate the
Is Available
checkbox.
6. Close the
Sample Holder Gallery
dialog.
7. Close the
Stage Navigation Settings
dialog.
8. From the
Panel Configuration Bar
, select
Variable Stage Bias
.
à
The
Variable Stage Bias
window is displayed.
9. Deactivate the
Beam deceleration carousel
checkbox.
10. Activate the
Stage bias low voltage
check-
box.
11. Use the
Stage bias Low Volts
scroll bar to adjust the voltage until you are satisfied with
the result.
90
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006