7 Troubleshooting | 7.2 Column
Procedure
1. Locate and unscrew the retaining screws using
a hex key 3 mm.
INFO:
Do not remove the screws as they will
be removed with the anode later.
2. Use long reach flexible tweezers to carefully re-
move the anode along with the retaining
screws.
3. Close the gun lid to prevent ingress of dust.
4. Put a small amount of polish on a piece of
printing paper.
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006
131