7 Troubleshooting | 7.2 Column
7. Check that
New Filament
is activated.
8. In the
Apertures
tab of the
SEM Controls
panel, click
Emission
.
à
The
Emission
button changes to
Normal
.
9
The filament starts running up.
INFO:
If you start up a new filament for the first couple of times, then the start-up may take a
bit of time. This is intended and does minimize damage to the new filament due to thermal
shock.
7.2.2.5.2 Aligning the Emission Image
To achieve an optimal alignment of the beam, you need to align the emission image to the center
of the image area.
Prerequisite
ü
The filament is started up.
Procedure
1. Switch on the crosshairs via
Menu Bar > View > Crosshairs
.
2. In the
SEM Controls
panel, select the
Gun
tab.
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006