ZEISS
2 Safety | 2.4 Warning Labels and Lights
Position and Figure of the Safety La-
bel
Description
9
At the gun head
Risk of burn
Before opening the gun leave the gun and fila-
ment cool down for 15 minutes after switching
off the microscope.
10
At the rear side of the plinth
Radiation hazard
Do not remove any parts.
Observe local safety and X-ray protection regula-
tions.
Service only through authorized personnel.
Reorder no. 347800-0033-000-06en
11
At the rear side of the plinth
Caution
X-rays are produced in this Scanning Electron Mi-
croscope!
The acceleration voltage is limited to 30 kV.
Observe local safety and X-ray protection regula-
tions.
Order no. 346000-0088-000
12
At the rear side of the plinth
Contact may cause burn or electric
shock. Only authorised service
Staff is allowed to service the
equipment.
Disconnect power before opening.
Hazardous voltage inside
WARNING
347800-0033-000-03en
Hazardous voltage inside
Contact may cause burn or electric shock. Only
authorized service Staff is allowed to service the
equipment.
Disconnect power before opening.
Reorder no. 347800-0033-000-03en
14
At the rear side of the plinth
WARNING
High leakage current
347800-0033-000-07en
Ensure proper grounding.
Do not operate the electron
microscope without a separate
ground connection.
High leakage current
Ensure proper grounding.
Do not operate the electron microscope without
a separate ground connection.
Reorder no. 347800-0033-000-07en
15
At the rear side of the plinth
Risk of electrical shock
Residual voltage at the mains plug after unplug-
ging.
Wait at least 5 s after unplugging the mains plug
before touching the pins of the mains plug.
Reorder no. 344700-0033-000-11en
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006