9 Technical Data and Conformity | 9.4 Performance Data and Specifications | EVO 15
ZEISS
Parameter
Value
CEE MALE PLUG 2P3W 4 h 20 A (yellow) or
for 230 V
AC
a CEE MALE PLUG 2P3W 6 h
16 A (blue).
The building installation should provide the
corresponding CEE socket 1/N/PE with de-
sired approvals of the country used: either for
120 V
AC
a CEE FEMALE RECEPTACLE 2P3W
4 h 20 A (yellow) or for 230 V
AC
a CEE FE-
MALE RECEPTACLE 2P3W 6 h 16 A (blue).
To avoid disturbance from other installed ma-
chines Carl Zeiss Microscopy recommends to
use a separate power connection to the main
distribution panel.
Power consumption
max. 3000 VA, dependent on accessories
Protective ground
High leakage currents are present in the mi-
croscope. Therefore, the microscope has to
be connected to a separate protective
ground.
An exclusive grounding connection to earth
must be provided as part of the building in-
stallation, i.e. a grounding screw terminal (Ø
8 mm) which is directly connected to the PE
of the FEMALE RECEPTACLE as short as possi-
ble.
This grounding connection must not be com-
mon to other electrical equipment.
A grounding wire AWG10 (≥ 5 m) is delivered
with the microscope. It serves as connection
between the grounding screw terminal (Ø
8 mm) and the microscope (Ø 6 mm).
Cross section
: Min. AWG10 (between
grounding screw terminal and PE of the CEE
FEMALE RECEPTACLE)
Circuit breaker type (at house installation)
16 A
Type C
Protection class
Class I
Momentary interruption
Less than a half cycle
160
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006