START-UP
Carl Zeiss
Attachment lamp VIS-LED for transmitted light
Axio Imager
62
430000-7344-001
M70-2-0020 e 06/2009
3.30.2
Adjusting the attachment lamp
VIS-LED
•
Switch on the add-on illuminator via the
microscope.
•
Switch off the diffusion disk for transmitted
light and remove the filter wheels for
transmitted light.
•
Use
≤
40x objective to focus on a free area of
the specimen.
•
Remove the eyepiece. Use the adjusting screws
(3-43/
1
and
2
) to center the image of the
illuminator in the pupil image.
•
3
) to optimize the
pupil image evenly.
•
Switch the diffusion disk for transmitted light
on again and insert the filter wheels for
transmitted light.
Fig. 3-43
Adjusting the attachment lamp
VIS-LED