OPERATION
Carl Zeiss
Operating the motorized microscope via the touchscreen of the TFT display
Axio Imager
130
430000-7344-001
M70-2-0020 e 06/2009
RL-illumination
The light source to be employed in reflected light
must be indicated here.
If a switching mirror for two illuminators is used,
this mirror will appear in this configuration menu
following successful initialization. Operation is
effected either via the function buttons (for
allocation, see below), or the function concerned is
automatically linked with the selection of an
allocated reflector module, so that the user does
not have to pay attention to any additional
components.
Sources of illumination:
−
None
−
HAL int.
−
HAL ext.
−
HBO
−
FluoArc
−
HXP
−
Colibri
−
LED int.
−
LED ext.
−
VisLED int.
−
VisLED ext.
The addition "internal" or "external" denotes that
the component is hooked up either to the power
supply incorporated in the stand or to an external
CAN bus power supply connected to the stand via
CAN bus.
Only types of lamps that are different can be
configured. An error message will appear in the
event of dual allocation.
Fig. 4-78
RL illumination
Fig. 4-79
RL illumination error message