OPERATION
Axio Imager
Operating the motorized microscope via the touchscreen of the TFT display
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
113
Working with two power supplies and two
illuminators:
Second-generation Axio Imagers can be equipped
with an additional external CAN bus enabled
power supply. When such a device is detected
during the switch-on operation, it will be
automatically initialized. In order for the power
supply to be used, a light source must be allocated
to it under
Settings -> Components -> Miscellaneous ->
TL/RL Illumination
.
Using two power supplies, two halogen or LED
illuminators may be operated in reflected and
transmitted light at their full power, completely
separately or jointly. The desired illuminator is
selected via the
On
or
Off
buttons on the TFT, the
light control buttons or, if there is no shutter in
transmitted light, even directly via the incremental
encoder of the light control.
Generating mixed light:
To generate mixed light, the two illuminators need
to be switched on. The two power supplies are
then controlled synchronously. This means that the
brightness of the two illuminators is adjusted up
and down in parallel, independently of the
intensity regulator.
(3)
F/A TL tab
The luminous-field diaphragm diameter and the
aperture diaphragm for the transmitted-light path
can be set here.
This function will be available only if a
motorized field diaphragm or aperture
diaphragm is installed.
Fig. 4-45
Using of two power supplies and
illuminators
Fig. 4-46
Microscope -> Light -> F/A TL