INSTRUMENT
DECRIPTION
Axio Imager
Eyepieces
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
27
When immersion objectives are used, the air between the cover slip and the objective is replaced by a
liquid, which in most cases is immersion oil. The plastic oiler containing 20 ml of Immersol 581 F
®
immersion oil (n
D
= 1.518) is particularly suitable for this purpose.
To prevent oil contamination of the specimen when the nosepiece is turned, the resilient mounts of the
immersion objectives can be locked in their lifted position by turning them clockwise (do not forget to
unlock them again!).
2.6
Eyepieces
The field-of-view number of the eyepieces PL 10x/25 Br. foc. and E-PL 10x/25 Br. foc. is 25 mm; that of
eyepieces W-PL 10x/23 Br. foc. and E-PL 10x/23 Br. foc. is 23 mm.
W-PL and PL in the eyepiece designation refers to the excellent image flatness up to the edge of the field
of view.
If required, eyecups for the eyepieces can be ordered under Cat. No. 444801-0000-000.
2.7
Stage micrometers and eyepiece reticles
Measuring and counting using a microscope requires stage micrometers and eyepiece reticles, a selection
of which is listed below:
Illustration
Designation, technical data
Cat. No.
Stage micrometer, positive 5 + 100/100 y
D = 0.17 mm
Graduation on +y-axis: 5 mm in 5 intervals
Graduation on –y-axis: 1 mm in 100 intervals
with 2 opposing scales = 10 µm, accuracy
±
1µm
474026-0000-000
Crossline micrometer 14:140/d = 26 mm
Graduation length = 14 mm
Increments = 0.1mm
Graduation tolerance
≤
0.001 mm
454060-0000-000