OPERATION
Axio Imager
Switching the HBO 100 on/off
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
87
4.4
Switching the HBO 100 on/off
•
The HBO 100 illuminator, which is used in place of the HAL 100 illuminator for fluorescence contrast
examinations, is to be switched on and off at the on/off switch (4-10/
2
) of the HBO 100 transformer.
•
After any illuminator switch off, allow for a cool down of approx. 15 minutes before switching on the
illuminator again. Failure to do so will unnecessarily shorten the service life of the mercury vapor short-
arc lamp.
4.5
Binocular phototube 30°/25 mot.
with two camera ports (2TV tube
mot.)
The 2TV tube mot. (4-11) is available for all
Axio Imager stands of type .M2 / .M2m and .Z2 /
.Z2m.
Like all other binocular tubes this tube is to be
mounted to the coupler plate and connected to
the stand via the CAN bus. To this end, connect
the cable to one of the three available CAN
connectors (Fig. 3-50).
The 2TV tube has two TV ports (4-11/
2
).
The front TV port is adjustable in the X, Y and Z
axis and to the rear (fixed) TV port.
The tube is configured via the TFT display or the
MTB 2004.
LEDs on the tube display (4-11/
3
the selected light path setting.
If you find the illumination of the display
disturbing, you can switch it off by pushing the
eyepiece shutter button (4-13/
1
) for three seconds.
Fig. 4-11
Binocular phototube 30°/25 mot.
with two camera ports (2TV tube
mot.)
Fig. 4-12
Tube display