Adjusting the transmitted-light DIC with
motorized four-position modulator
turret for transmitted light
•
On the nosepiece, swivel in the objective
suitable for DIC.
•
On the reflector turret, swivel in analyzer
module DIC ACR shiftfree for transmitted light.
•
On the modulator turret, swivel in the
appropriate DIC prism by
pressing
on knob
1
) for forward rotation or on knob
2
) for backward rotation of the turret.
•
Adjust the luminous-field diaphragm and
aperture diaphragm according to KÖHLER.
•
Set optimum contrast by turning knob
(4-121/
2
).
•
If desired, compensator
λ
can be inserted into
the slot for the analyzer slider located above the
nosepiece, so as to create a differential
interference contrast (DIC) in colours.
Fig. 4-121 Motorized four-position modulator