START-UP
Carl Zeiss
Screwing in objectives
Axio Imager
38
430000-7344-001
M70-2-0020 e 06/2009
3.8
Screwing in objectives
•
Move the mechanical stage with stage carrier to
the lower stop position.
•
2
) from the respective
openings on the objective nosepiece.
•
1
) from the case and
screw them into the nosepiece (3-9/
3
) starting
with the lowest magnification (clockwise
rotation).
•
Make sure to cover objective mounts of the
nosepiece not in use by dust protection caps.
In combination with objectives with
W
0.8" thread please use adapter
M27x0.75 "0" on W 0.8 (000000-
1095-168).
3.9
Attaching or changing the
condenser
•
Turn coarse focusing drive to move the stage
carrier to the upper stop position.
•
Turn height control (3-10/
1
) to move condenser
carrier (3-10/
2
) down.
•
Slightly unscrew locking screw (3-10/
4
)
(SW 1.5), if necessary.
•
If available, swivel out front lens (3-10/
8
) of
condenser by means of lever (3-10/
7
).
•
Directing the locking screw on the bottom of
the condenser to the groove (3-10/
3
), insert the
6
) between condenser carrier
2
9
) in the
centerable condenser mount (3-10/
5
) as far as it
will go.
•
Tighten locking screw (3-10/
4
) on condenser
mount. In doing so, do not apply excessive
force to avoid any damage to the condenser
mount.
Fig. 3-9
Screwing in objectives
Fig. 3-10
Attaching the achromatic-aplanatic
universal condenser