Replacing the HAL
100 halogen
lamp
CAUTION
Hot surface!
You need not remove the lamp
housing from the stand to replace the
halogen lamp.
Do not
store the
supplied replacement tool (3-39/
7
) for
lamp replacement in the lamp housing
while the illuminator is operating.
The replacement lamp (3-39/
8
) may
remain put on in the lamp housing.
•
Switch off the microscope as described in
Section 4.3, disconnect plug (3-35/
7
) from
connector 12 V/100 W (3-35/
3
– reflected light
or
5
– transmitted light) and allow for a
cool down of approximately 15 minutes.
•
3
) of halogen
illuminator HAL 100 (3-39/
1
), fully pull out lamp
carrier (3-39/
2
) and put it down separately.
•
Depress both spring levers (3-39/
5
) and remove
the old halogen lamp (3-39/
6
) upward.
•
Depressing both spring levers insert the new
lamp in the lamp socket (3-39/
4
) and release
the spring levers. Always hold/grasp the
halogen lamp by means of the replacement tool
(3-39/
7
), as even traces of grease on the lamp
may affect its lifetime.
•
Shortly depress spring levers once more to
center the lamp.
•
Insert lamp carrier until it locks.
Fig. 3-39
Changing the halogen lamp