<8. Detailed Data Setting>
8-11
IM 11M13A01-04E
8.6 Contact Input Setting
The analyzer contact inputs execute set functions by accepting a remote (contact) signal. Table 8.12
shows the functions executed by a remote contact signal.
Table 8.12
Contact Input Functions
Set item
Function
Calibration gas pressure
decrease
While a contact signal is on, neither semi-automatic nor automatic calibrations
can be made.
Measuring range change
While contact input is on, the analog output range is switched to 0-25%O
2
.
Calibration start
If a contact signal is applied, semi-automatic calibration starts (only if the semi-
automatic or automatic mode has been setup). Contact signal must be applied
for at least one second. Even though a continuous contact signal is applied, a
second calibration cannot be made. If you want to make a second calibration,
turn the contact signal off and then back on.
Unburnt gas detection
If a contact signal is on, the heater power will be switched off. (An one-to
11-second time interval single-output signal is available as a contact signal.)
If this operation starts, the sensor temperature decreases and an error occurs.
To restore it to normal, turn the power off and then back on, or reset the analyzer.
CAUTION
• To conduct a semi-automatic calibration, be sure to set the Calibration setup mode to
“Semi-automatic” or “Automatic”.
8.6.1 Setting Contact Input
To set the contact inputs, follow the parameter codes given in Table 8.13.
Table 8.13
Parameter Codes for Contact Input Settings
Set item
Parameter code
Set value
Contact input 1 (function)
E01
0
Invalid
1
Calibration gas pressure
decrease
2
Measuring range change
3
Calibration
4
Unburnt gas detection
Contact input 2 (function)
E02
0
Invalid
1
Calibration gas pressure
decrease
2
Measuring range change
3
Calibration
4
Unburnt gas detection
Contact input 1 (action)
E03
0
Operated when closed
1
Operated when open
Contact input 2 (action)
E04
0
Operated when closed
1
Operated when open
8.6.2 Default Values
When the analyzer is delivered, or if data are initialized, the contact inputs are all open.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......