<7. Startup>
7-8
IM 11M13A01-04E
7.5 Confirmation of Equipment Type Setting
This equipment can be used for both the Oxygen Analyzer and the Humidity Analyzer. If you choose
optional specification /HS at the time of purchase, the equipment is set for the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired model has been set.
Note that if the equipment type setting is changed after operating data are set, the operating data
that have been set are then initialized and the default settings remain. Set the equipment type with
parameter code [F01]. See Table 10.7, later in this manual.
CAUTION
Note that if the equipment type is changed, operation data that have already been set are initialized
(reverting to the default setting).
Table 7.5
Equipment Type Setting Procedure
Switch operation
Display
Description
>
∧
ENT
A
01
Display after the password has been entered.
>
∧
ENT
F
01
Touch the [
∧
] key to switch to Group F. If an unwanted alphabetic character after F
has been entered, continuously touch the [
∧
] key to return to the original.
>
∧
ENT
0
Touch the [ENT] key for confirmation. If 0 (zero) is entered, the oxygen analyzer is
already set. If 1 (one) is entered, the humidity analyzer has been set.
Change the setting following the steps below.
>
∧
ENT
0
Continuously touch the [
∧
] key, and the position of the digit will change from 1 to 0
to 1 to 0. Release the [ENT] key when 0 is displayed.
>
∧
ENT
0
Touch the [ENT] key. The numeric value will flash.
>
∧
ENT
0
Touch the [ENT] key again to stop the numeric value from flashing.
>
∧
ENT
F
01
Touch the [ENT] key once again, and the display will change to the parameter
code.
>
∧
ENT
Basic
panel
display
Touch the [
>
] key together with the [ENT] key to return to the basic panel display.
(This is not required if you proceed to make another setting.) (The displayed
numeric values indicate the measurement gas concentration.)
The symbol [
] indicates that the key is being touched. Light characters indicates that the digits are flashing.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......