<7. Startup>
7-10
IM 11M13A01-04E
Table 7.7
Minimum and Maximum Current Setting Procedure
Switch operation
Display
Description
>
∧
ENT
A
01
Display after the password has been entered.
>
∧
ENT
C
01
Set the oxygen concentration at 4 mA. Change the parameter code to C11.
Touch the [
∧
] key to switch to Group C.
>
∧
ENT
C
0
1
Touch the [
>
] key to move the position of the digit that is flashing to the right.
>
∧
ENT
C
1
1
Touch the [
∧
] key to enter the numeric value to 1.
>
∧
ENT
0
00
Touch the [ENT] key to display the current set value (0% O
2
has been set).
>
∧
ENT
0
0
0
Touch the [
>
] key to move the position of the digit that is flashing to the right.
>
∧
ENT
0
1
0
Touch the [
∧
] key to change the numeric value to 1.
>
∧
ENT
010
If you touch the [ENT] key, all the digits flash.
>
∧
ENT
010
Touch the [ENT] key again to stop the flashing.
>
∧
ENT
C
11
Touch the [ENT] key once again, and the display will switch to the parameter code
selection display.
>
∧
ENT
C1
1
Set the oxygen concentration at 20 mA. Touch the [
>
] key to move the position of
the digit that is flashing to the right.
>
∧
ENT
C1
2
Touch the [
∧
] key to enter the numeric value to 2.
>
∧
ENT
0
25
Touch the [ENT] key to display the current set value.
>
∧
ENT
02
5
Touch the [
>
] key to move the position of the digit that is flashing to the right.
>
∧
ENT
02
0
Touch the [
∧
] key to change the numeric value to 0. The numeric value will change
from 5 to 6 ... to 9 and back to 0.
>
∧
ENT
020
If you touch the [ENT] key, all the digits flash.
>
∧
ENT
020
Touch the [ENT] key again to stop the flashing.
>
∧
ENT
C
12
Touch the [ENT] key once again to switch to the parameter code selection display.
>
∧
ENT
Basic
panel
display
Touch the [
>
] key together with the [ENT] key to return to the basic panel display.
(This is not required if you proceed to make another setting.)(The displayed
numeric values indicate the measurement gas concentration.)
The symbol [
] indicates that the key is being touched. Light characters indicates that the digits are flashing.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......