<3. Installation>
3-4
IM 11M13A01-04E
Note 3: Operation
• Keep the “WARNING” label to the Analyzer.
WARNING:
OPEN CIRCUIT BEFORE REMOVING COVER. INSTALL IN
ACCORDANCE WITH THE INSTRUCTION MANUAL
IM 11M13A01–04E. USE AT LEAST 80°C HEAT
RESISTANT CABLES.
• Take care not to generate mechanical sparking when accessing to the analyzer and peripheral
devices in hazardous area.
Note 4: Maintenance and Repair
• The analyzer modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void Factory Mutual Explosion-proof
Approval.
3.1.4 CSA Explosion-proof Type
ZR202S–C Analyzer for use in hazardous area:
Note 1: Applicable Standard: C22.2 No.0-M1991, C22.2 No.0.4-04, C22.2 No.0.5-1982,
C22.2 No.25-1966, C22.2 No.30-M1986, C22.2 No.94-M91,
C22.2-No.61010-1-04
Certificate Number:
1649642
Type of protection:
Explosion-proof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure:
Type 4X
Temperature Class:
T2
Note 2: Wiring
• All wiring shall comply with Canadian Electrical Code Part 1 and Local Electrical Code.
• In hazardous area, wiring shall be in conduits as shown in the figure.
WARNING:
SEAL ALL CONDUITS WITHIN 50 cm OF THE ENCLOSURE.
UN SELLE DOIT ÊTRE INSTALLÊ Á MOINS DE 50 cm DU BîTIER.
Hazardous Area Division 1
50cm MAX.
Non-hazardous
Area
Non-hazardous
Area
Conduit
Sealing Fitting
F3-2E.ai
50cm MAX.
Figure 3.3
Wiring of ZR202S-C
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......