<7. Startup>
7-14
IM 11M13A01-04E
7.9.3 Checking Contact Input
Follow Table 7.12 to check the contact input. The table uses an example with contact input 1.
Table 7.12
Checking Contact Input
Switch operation
Display
Description
>
∧
ENT
A
01
Display after the password has been entered.
>
∧
ENT
G
01
Touch the [
∧
] key to switch to Group G.
>
∧
ENT
G
0
1
Touch the [
>
] key to move the position of the digit that is flashing to the right one
digit.
>
∧
ENT
G
2
1
Touch the [
∧
] key to enter 2.
>
∧
ENT
0
Touch the [ENT] key. 0 is displayed with the contact input open. If the contact input
is closed, the display will be 1 (one).
This enables you to check whether or not the wiring connections have been
properly made or not.
>
∧
ENT
G
21
Touch the [ENT] key once again to switch the parameter code selection display.
>
∧
ENT
Basic
panel
display
Touch the [
>
] key together with the [ENT] key to return to the basic panel display.
The symbol [
] indicates that the key is being touched. Light characters indicates that the digits are flashing.