<5. Wiring>
5-1
IM 11M13A01-04E
5. Wiring
This chapter describes wiring procedures necessary for the EXAxt ZR Integrated type Explosion-proof
Zirconia Oxygen Analyzer.
5.1 General
CAUTION
Be sure to read Subsections 3.1.2 to 3.1.5 where the important information on wiring is provided.
CAUTION
• Never supply current to the equipment or any other device constituting a power circuit in
combination with the equipment, until all wiring is completed.
• This product complies with CE marking.
Where a performance suit for CE marking is necessary, the following wiring procedure is necessary.
1. Install an external switch or circuit breaker to the power supply of the equipment.
2. Use an external switch or circuit breaker rated 5 A and conforms to IEC 947-1 or IEC 947-3.
3. It is recommended that the external switch or circuit breaker be mounted in the same room as
the equipment.
4. The external switch or circuit breaker should be installed within the reach of the operator, and
marked as the power supply switch of this equipment.
Wiring procedure
Wiring should be made according to the following procedure:
1. Be sure to connect the shield of the shielded line to FG terminal of the analyzer.
2. The most outer sheath of the signal line and the power cable should be stripped off to the
minimum necessary length.
3. Signal will be affected by noise emission when the signal lines, power cable and heater cable
are located in the same conduit. When using a conduit, signal lines should be installed in the
separate conduit from power and heater cables. Be sure to ground the metal conduit.
4. Mount the attached two blind plugs to unused cable connection gland(s) of the equipment.
5. The cables indicated in Table 5.1 are used for wiring.
6. After completing the wiring, screw the cover in the terminal box body and secure it with a lock screw.
Table 5.1 Cable Specifications
Terminal name of analyzer
Name
Need for shields
Number of wires
L, N,
Power supply
2 or 3 *
AO+, AO-
Analog output
Ο
2
DO-1, DO-2
Contact output
2 to 8
DI-1, DI-2, DI-C
Contact input
3
Note *: When the case is used for protective grounding, use a 2-wire cable.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......