<10. Other Functions>
10-13
IM 11M13A01-04E
Calibration-related Items in Group B
Code
Item
Tuning
Engineering unit
Default setting
B01 Zero gas concentration
0.3 to 100
%O
2
1%O
2
B02 Span gas concentration
4.5 to 100
%O
2
21%O
2
B03 Calibration mode
0 Manual calibration
Manual calibration
1 Semi-automatic and manual
calibration
2 Automatic, semi-automatic, and
manual calibration
B04 Hold (Output stabilization)
time
0 minutes, 0 seconds to 60 minutes,
59 seconds
MM.SS
10 minutes,
0 seconds
B05 Calibration time
0 minutes, 0 seconds to 60 minutes,
59 seconds
MM.SS
10 minutes,
0 seconds
B06 Calibration interval
0 days 0 hours to 255 days 23 hours
DD.HH
30 days, 0 hours
B07 Calibration start date and
time
YY.MM.DD
HH.MM
00.01.01.00.00
B08 Calibration procedure
0 Zero and span
Zero and span
1 Span only
2 Zero only
B09 Calibration concentration
measurement
Display only
% O
2
B10 Manual calibration
implementation
B11 Semi-automatic calibration
implementation
Continue to the next page