<10. Other Functions>
10-17
IM 11M13A01-04E
10.5 Handling of the ZO21S Standard Gas Unit
WARNING
Use only non-hazardous area.
The following describes how to flow zero and span gases using the ZO21S Standard Gas Unit.
Operate the ZO21S Standard Gas Unit, for calibrating a system classified as System 1, according to
the procedures that follow.
10.5.1 Standard Gas Unit Component Identification
Applies the power to operate the
pump to supply the span gas.
Carrying case
Flow checker
Span gas valve
Checks the zero and
span gas flow.
Controls the span gas (air) flow.
Zero gas valve regulator
Tube connection
Cover screws (six pcs.)
Pump
Supplies span gas (air)
Gas cylinder
Contains the zero gas. A gas
of 7 Nl is charged to 700 kPa
Clamp
Clamps the gas cylinder.
Zero gas valve
Attaches to the gas
cylinder for use.
Power cord
F10-4E.ai
Figure 10.4
Standard Gas Unit Component Identification