<12. Troubleshooting>
12-2
IM 11M13A01-04E
12.1.2 Remedies When an Error Occurs
Err-01: Cell Voltage Failure
Err-01 occurs when the cell (sensor) voltage input to the converter falls below -50 mV (corresponding
to about 200 vol%O
2
). The following are considered to be the causes for the cell voltage falling below
-50 mV:
(1) Continuity failure between the sensor assembly electrode and the contact.
(2) Damage or deterioration of the sensor assembly.
(3) Improper connection between the sensor and the electronics.
(4) Wiring failure inside the detector.
(5) Abnormality in electrical circuits inside the converter.
<Locating cause of failure, and countermeasures>
1) Turn off the power to the equipment.
2) Remove the sensor assembly from the probe. Check for dirty or corroded sensor parts, including
electrode and contact.
3) If the contact part is normal, the sensor assembly may be damaged or deteriorated. Replace the
sensor assembly. In this case, be sure to replace the metal O-ring and contact.
4) If Err-01 still occurs, check that the sensor and the electronics are properly connected.
5) Remove the probe to gain access to the two connectors (four connectors for the optional
automatic calibration unit), as indicated in Figure 12.2. Check these connectors are properly
connected.
6) If Err-01 still occurs, the electronics may be defective. Contact your local Yokogawa service or
sales representative.
Err-02: Heater Temperature Failure
This error occurs if the detector heater temperature does not rise during warm-up, or if the
temperature falls below 730°C or exceeds 780°C after warm-up is completed.
Causes considered for cases where Err-02 occurs independently are shown below.
(1) Faulty heater in the detector (heater wire breakage).
(2) Faulty thermocouple in the detector.
(3) Failure in electrical circuits inside the converter.
<Locating cause of failure, and countermeasures>
(1) Turn off the power to the analyzer.
(2) Remove the probe from the analyzer. Also remove all the connectors between the converter
and probe. Measure the resistance of the heater wire (yellow wire) from the probe as indicated
in Figure 12.2. The heater assembly is normal if the resistance is lower than about 90 Ω. If the
resistance is higher than that value, the heater assembly may be defective. In this case, replace
the heater assembly (refer to Subsection 11.1.3, “Replacement of the Heater Assembly”).
F12-2E.ai
Multimeter
(Ω)
Heater wire
Figure 12.2
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......