<8. Detailed Data Setting>
8-10
IM 11M13A01-04E
WARNING
• Contact output 2 is linked to the detector’s heater power safety switch. As such, if contact
output 2 is on, the heater power stops and an Err-01 (cell voltage abnormal) or Err-02 (heater
temperature abnormal) occurs.
8.5.3 Default Values
When the analyzer is delivered, or if data are initialized, contact outputs are by default as shown in
Table 8.11.
Table 8.11
Contact Output Default Settings
Item
Contact output 1 Contact output 2
High-high alarm
High alarm
Low alarm
Low-low alarm
Error
O
During warm-up
O
Output range change
During calibration
During maintenance
O
Calibration gas pressure decrease
Unburnt gas detection
Operating contact status
Open
Closed (fixed)
O: Present
NOTE
The above blank boxes indicate the items have been set off.