
<3. Installation>
3-5
IM 11M13A01-04E
Note 3: Operation
• Keep the “WARNING” label to the Analyzer.
WARNING:
OPEN CIRCUIT BEFORE REMOVING COVER. REFER TO
IM 11M13A01–04E. USE AT LEAST 80°C HEAT RESISTANT CABLES.
OUVRIR LE CIRCUIT AVANT D’ENLEVER LE COUVERCLE. UTILISEZ DES CÁBLES
RÊSISTANTS Á 80°C MINIMUM. VEUILLEZ VOUS RÊFÊRER AU IM 11M13A01–04E.
• Take care not to generate mechanical sparking when accessing to the analyzer and peripheral
devices in hazardous area.
Note 4: Maintenance and Repair
• The analyzer modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void Canadian Standards Explosion-proof
Certification.
3.1.5 IECEx Flameproof Type
ZR202S–D Analyzer for use in hazardous area:
Note 1: IECEx flameproof type
Applicable Standard: IEC 60079-0: 2011, IEC 60079-1: 2014, IEC 60079-31: 2013
Certificate Number:
IECEx KEM 06.0006X
The symbol “X” placed after certificate number indicates that the equipment is subjected to
special conditions for safe use. Refer to Note 6.
Type of protection:
Ex db IIB + H2 Gb,
Ex tb IIIC T300ºC Db
Temperature class for Ex “db”: T2
The maximum surface temperature for Ex “tb”: T300ºC
Degree of protection of enclosure: IP66
Note 2: Wiring
• All wiring shall comply with local installation requirement.
Note 3: Operation
• Keep to “WARNING” on the Analyzer.
WARNING:
DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE IS PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD-READ USER’S
MANUAL (IM11M13A01-04)
FOR INSTALLATION AND SAFE USE, READ IM11M13A01-04
USE AT LEAST 80ºC HEAT RESISTANT CABLES & CABLE GLANDS
• Take care not to generate mechanical sparking when accessing to the analyzer and peripheral
devices in hazardous area.
Note 4: Maintenance and Repair
• The analyzer modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......