<7. Startup>
7-2
IM 11M13A01-04E
7.1 Checking Piping and Wiring Connections
Refer to Chapters 4 and 5, earlier in this manual, for piping and wiring confirmations.
7.2 Valve Setup
Set up valves and associated components used in the analyzer system in the following procedures:
(1) If a stop valve is used in the calibration gas inlet, fully close this valve.
(2) If instrument air is used as the reference gas, adjust the Air set secondary pressure so that
the air pressure of sample gas pressure plus approx. 50 kPa (plus approx. 150 kPa for with
check valve) (300 kPa maximum) is obtained. Turn the reference gas flow setting valve in
the ZA8F flow setting unit to obtain the flow of 800 to 1000 mL/min. (Turning the valve shaft
counterclockwise increases the rate of flow. When turning the valve shaft, if the valve has a lock
nut, first loosen the lock nut.) After completing the valve setup, be sure to tighten the lock nut.
NOTE
The calibration gas flow setting will be described later. Fully close the needle valve in the flow setting
unit.
7.3 Supplying Power to Analyzer
CAUTION
To avoid temperature changes around the sensor, it is recommended that the power be continuously
supplied to the Oxygen Analyzer if it is used in an application where its operations and suspensions
are periodically repeated.
It is also recommended to flow a span gas (instrument air) beforehand.
Supply power to the analyzer. A display as in Figure 7.2, which indicates the detector’s sensor
temperature, then appears. As the heat in the sensor increases, the temperature gradually rises
to 750°C. This takes about 20 minutes after the power is turned on, depending somewhat on the
ambient temperature and the sample gas temperature. After the sensor temperature has stabilized
at 750°C, the analyzer is in the measurement mode. The display panel then displays the oxygen
concentration as in Figure 7.3. This is called the basic panel display.
F7-2E.ai
%
F7-3E.ai
Figure 7.2 Display of Sensor Temperature
Figure 7.3 Measurement Mode Display
While Warming Up
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......