<9. Calibration>
9-7
IM 11M13A01-04E
Setting
When setting calibration timing requirements, bear the following precautions in mind:
NOTE
(1) If the calibration interval is shorter than the sum of hold (output stabilization) time plus calibration
time, the second calibration start time will conflict with the first calibration. In such a case, the
second calibration will not be conducted. (When both zero and span calibrations are to be
performed, the calibration time is double that required for a single (zero or span) calibration.)
(2) For the same reason, if the calibration start time conflicts with manual calibration or semi-
automatic calibration, the current calibration will not be conducted.
(3) If the calibration time conflicts with maintenance service operation, calibration will start after
completing the maintenance service operation (see Subsection 8.3.1, earlier in this manual).
(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration will be
conducted; a second or later calibration will not be conducted.
(5) If a past date is set to the calibration start day, no calibration will be conducted.
Table 9.1
Parameter Codes for Calibration Setting
Set Item
Parameter
code
Set value
Engineering
unit
Zero gas concentration
B01
Set Zero gas concentration
%O
2
Span gas concentration
B02
Set Span gas concentration
%O
2
Calibration mode
B03
0 Manual calibration
1 Semi-automatic and manual
2 Automatic, semi-automatic, and manual
Hold (Output stabilization) time
B04
0 minutes 0 seconds to 60 minutes 59 seconds MM.SS
Calibration time
B05
0 minutes 0 seconds to 60 minutes 59 seconds MM.SS
Calibration interval
B06
0 days 0 hours to 255 days 23 hours
Date and time
Start date and time
B07
Date and time of first calibration
YY.MM.
DD.HH.MM
Calibration procedure
B08
0 : Zero and span
1 : Span only
2 : Zero only
Default Values
When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as
shown in Table 9.2.
Table 9.2
Default Settings for Calibration
Item
Default Setting
Calibration mode
Manual
Calibration procedure
Span - zero
Zero gas (oxygen) concentration
1.00%
Span gas (oxygen) concentration
21.00%
Hold (Output stabilization) time
10 minutes, 00 seconds
Calibration time
10 minutes, 00 seconds
Calibration interval
30 days, 00 hours
Start date and time
00 (YY) 01 (MM) 01(DD) 00:00
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......