<7. Startup>
7-5
IM 11M13A01-04E
Table 7.2
Display Functions
Display
Function and item to be set
Basic panel
Displays the oxygen concentration in normal operation, or displays the detector
heater temperature while warming up. If an error or alarm arises, the corresponding
error or alarm number appears.
Password entry
Enters the password for the parameter code selection display.
Group A setup
Displays detailed data, such as the cell voltage or temperature.
Group B setup
Sets and performs calibration.
Group C setup
Sets analog output.
Group D setup
Sets an alarm.
Group E setup
Sets the contact inputs and contact outputs.
Group F setup
Selects the type of equipment and sets the parameters for computation.
Group G setup
Performs the current-loop or contact checks.
7.4.3 Entering Parameter Code Selection Display
This subsection briefly describes the password entry procedure for entering the parameter code
selection display. The password is 1102 - it cannot be changed to a different password.
Table7.3
Parameter Code Selection
Switch operation
Display
Description
>
∧
ENT
21.0%
Warm-up is complete, and the basic panel is now displayed.
>
∧
ENT
PASSno
Continuously touch the [ENT] key for at least three seconds
to display “PASSno”.
>
∧
ENT
0
000
Touch the [ENT] key again. This allows you to change the leftmost
digit that is flashing.
>
∧
ENT
1
000
Set the password 1102. If you touch the [
∧
] key, the digit that is flashing will
be 1.
>
∧
ENT
1
0
00
Touch the [
>
] key to move the position of the digit that is flashing to the
right one digit.
>
∧
ENT
1
1
00
Touch the [
∧
] key to change the numeric value to 1.
>
∧
ENT
110
0
Touch the [
>
] key again to move the position of the digit that is flashing to
the right one more digit. Continuously touch the [
>
] key, and the position of
the digit that is flashing will move continuously to the right.
>
∧
ENT
110
2
Touch the [
∧
] key to change the numeric value to 2. Continuously touch [
∧
]
key, and the numeric value increases continuously.
>
∧
ENT
1102
If you touch the [ENT] key, all the digits flash.
>
∧
ENT
A
01
Touch the [ENT] key again to display A01 on the parameter code selection
display.
The symbol [
] indicates that the key is being touched. Light characters indicate that the digits are flashing.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......