
<2. Specifications>
2-5
IM 11M13A01-04E
Instrument Air System:
Pressure;
50 kPa + the pressure inside the furnace
150 kPa + the pressure inside the furnace with automatic calibration unit or
check valve. (It is recommended to use air which has been dehumidified by
cooling to dew point -20°C or less, and dust or oil mist are removed.)
Consumption; Approx. 1.5 Nl/min
Oxygen concentration of calibration gas:
For explosion-proof use, not more than that found in normal air, typically 21 vol%
Wetted Material:
316 SS (JIS), Zirconia, 304 SS (JIS) (flange),
Hastelloy B, (Inconel 600, 601)
Construction: Heater and thermocouple replaceable construction. Equivalent to NEMA 4X/IP66
(Achieved when pipes are installed at calibration gas and reference gas inlets and
exhaust pipe is installed so that reference gas can be exhausted to clean
atmosphere. Excluding probe top.)
(Achieved when the cable entry is completely sealed with a cable gland.)
Gas Connection:
Rc 1/4 or 1/4 NPT(F)
Wiring Connection:
ATEX: M20 by 1.5 mm, 1/2 NPT select one type (4 pieces)
FM:
1/2 NPT (4 pieces)
CSA: 1/2 NPT (4 pieces)
IECEx: M20 by 1.5 mm, 1/2 NPT select one type (4 pieces)
Installation:
Flange mounting
Probe Mounting Angle: Horizontal to vertically downward.
Case:
Aluminum alloy
Finish:
Polyurethane corrosion-resistance coating
Weight:
Insertion length of 0.4m: approx. 15 kg (ANSI 150 4)
Insertion length of 1.0m: approx. 17 kg (ANSI 150 4)
Insertion length of 1.5m: approx. 19 kg (ANSI 150 4)
Insertion length of 2.0m: approx. 21 kg (ANSI 150 4)
Functions:
Display Function:
Displays values of the measured oxygen concentration, etc.
Alarm, Error Display:
Displays alarms such as “AL-06” or errors such as “Err-01” when any such
status occurs.
Calibration Functions:
Automatic Calibration; Requires the Automatic Calibration Unit. It calibrates
automatically at specified intervals.
Semi-automatic Calibration; Requires the Automatic Calibration Unit. Input
calibration start signal by optical switch or contact, then it calibrates
automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration
gas in operation interactively with the optical switch.
Maintenance Functions: Can operate updated data settings in daily operation and checking.
Display data settings, calibration data settings, test settings (current output
loop check, contact input/output check).
Setup Functions:
Initial settings suit for the plant conditions when installing the analyzer.
Current output data settings, alarm data settings, contact data settings, other
settings.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......