<10. Other Functions>
10-6
IM 11M13A01-04E
10.1.12 Heater On-Time Ratio
The probe sensor is heated to and maintained at 750°C. When the sample gas temperature is high,
the amount of heater ON-time decreases.
10.1.13 Oxygen Concentration (with time constant)
When the output damping is specified in the mA-output range setting, the corresponding time constant
is also displayed.
10.1.14 Maximum Oxygen Concentration
The maximum oxygen concentration and the time of its occurrence during the period specified in the
Averaging display are displayed. If the setup period elapses, the maximum oxygen concentration that
has been displayed so far will be cleared and a new maximum oxygen concentration will be displayed.
If the setup period of time is changed, the current maximum oxygen concentration will be displayed
(for more details, see Subsection 8.7.2 earlier in this manual).
10.1.15 Minimum Oxygen Concentration
The minimum oxygen concentration and the time of its occurrence during the period specified in the
Averaging display are displayed. If the setup period elapses, the minimum oxygen concentration that
has been displayed so far will be cleared and a new minimum oxygen concentration will be displayed.
If the setup period of time is changed, the current minimum oxygen concentration will be displayed (for
more details, see Subsection 8.7.2 earlier in this manual).
10.1.16 Average Oxygen Concentration
The average oxygen concentration during the periods over which average values are calculated is
displayed. If the setup period elapses, the average oxygen concentration that has been displayed so
far will be cleared and a new average oxygen concentration will be displayed. If the setup period of
time is changed, the current average oxygen concentration will be displayed (for more details, see
Subsection 8.7.2 earlier in this manual).
10.1.17 Span and Zero Correction Ratios
Span and zero correction ratios for the past ten calibrations are recorded to enable you to check the
degradation of the sensor (cell). If the correction ratio is beyond the limits as shown in Figure 10.2, the
sensor should no longer be used.
These ratios can be found by calculating the data as shown below.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......