7-8
IM 704210-01E
• Selecting the window size (number of windows)
Up to 16 windows can be set if the multi-window function is ON.
Setting range: 1 to 16
Setting the X-axis and Y-axis
For the setting method for center value of X-axis, span of X-axis, Y-axis scale, maximum value of
Y-axis and area, refer to 7.1.
If MULTI is selected, use the W.SPACE touch key of the X-axis/Y-axis menu to set the space
between the center value of X-axis for a window and that for other windows. The W.SPACE
touch key is displayed if AUTO is selected as the multi-window mode.
Setting range for the space between the center values of X-axis: 0 to 250 ns
Note
If AUTO is selected as the multi-window mode, the histogram containing the smallest measured data must
be used to set the X-axis and Y-axis for window 1. For the space between the center values of X-axis, only
a positive displacement is set. Thus, if a histogram contains measured data smaller than that contained in
the histogram used for window 1, no window will be assigned to that histogram. As a result, this histogram
will not be displayed even if the window is switched to another at steps 24 and 25.
7.2 Using the Multi-window Function