![YOKOGAWA TA320 User Manual Download Page 130](http://html.mh-extra.com/html/yokogawa/ta320/ta320_user-manual_911358130.webp)
11-18
IM 704210-01E
Testing method
• Wait for 30 minutes after the TA320 is turned ON.
• Set the frequency of the synthesized signal generator (8657B) to the values given in the table
below, and check that the measured values (mean value and standard deviation) are within the
allowable range using the time interval measurement function of the TA320.
Test results
•
A
B
Frequency (8657B) Avarage Value
Standard Deviation
Measured Value
Allowable Range
Measured Value
Allowable Range
10MHz
98.4 ns to 101.6 ns
320ps or less
20MHz
48.7 ns to 51.3 ns
310ps or less
•
A
B
Frequency (8657B) Avarage Value
Standard Deviation
Measured Value
Allowable Range
Measured Value
Allowable Range
10MHz
98.4 ns to 101.6 ns
320ps or less
20MHz
48.7 ns to 51.3 ns
310ps or less
•
A
B
Frequency (8657B) Avarage Value
Standard Deviation
Measured Value
Allowable Range
Measured Value
Allowable Range
10MHz
48.4 ns to 51.3 ns
320ps or less
20MHz
23.7 ns to 26.3 ns
310ps or less
•
A
B
Frequency (8657B) Avarage Value
Standard Deviation
Measured Value
Allowable Range
Measured Value
Allowable Range
10MHz
48.4 ns to 51.3 ns
320ps or less
20MHz
23.7 ns to 26.3 ns
310ps or less
11.6 Carrying out the Performance Test