5-4
IM 704210-01E
5.3 Setting Time Interval Measurement
≡
≡
≡
≡
≡
For a description of the function, refer to 1.3
≡
≡
≡
≡
≡
P r o c e d u r e
• To exit from the current menu, press any of the top menu keys or press the ESC touch key
displayed on the screen.
FUNCTION Top Menu
MODE
FUNCTION
DISPLAY
SAMPLE
INPUT
UTILITY
Displays the measurement function menu.
Selects the pattern.
Selects the slope (edge).
1.
Press the FUNCTION key to display the FUNCTION top menu.
Selecting time interval measurement
2.
Press the FUNC touch key to display the measurement function menu.
3.
Press the TIME INTERVAL touch key to select time interval measurement. The screen will
return to the FUNCTION top menu when the TIME INTERVAL touch key is pressed.
Selecting the pattern
4.
Press the CHA/CHB touch key to select the desired pattern (CHA-B or CHA-B-A). The
highlighted pattern (CHA-B or CHA-B-A) will be selected.
Selecting the slope (edge)
5.
Press the SLOPE touch key to highlight SLOPE.
6.
Turn the rotary knob to select the desired slope type.
Measurement function menu
Selects period measurement.
These keys are displayed for time stamp mode only.
Explanation
Measurement range
Measurement range varies according to the selected sampling mode and pattern.
• Time stamp mode:
• CHA-B pattern : 5 ns to 100 ms
• CHA-B-A pattern : 30 ns to 100 ms (for both A to B and B to A)
• Hardware histogram mode:
• CHA-B pattern : 5 ns to 3.2
µ
s
• CHA-B-A pattern : 30 ns to 3.2
µ
s (for both A to B and B to A)